40G Certifier Remote Calibration Data Missing

Document created by viavisupport on Feb 21, 2017
Version 1Show Document
  • View in full screen mode
Viavi 40G Certifier Remote Calibration Data Missing from the report.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           

Problem Overview

The remote unit calibration date is intermittently missing from report data for some test results on my 40G Certifier.

Resolution

The most likely cause of the calibration date being missing in the report data is procedural. The SAVE must be performed and completed prior to breaking the physical path between the Local and the Remote unit. If a cable or fiber is disconnected prior to the save completing then the calibration date of the remote unit will be missing. You should wait for the SAVE to complete before moving any cables/fibers from either the Local or the Remote units.